| Type | Automatic Potentiometric Titrator |
| Model | AT-710M | AT-710S |
| Product configuration | MCU-710M + AT-710 + IDP-100 + Propeller stirrer | MCU-710S + AT-710 + IDP-100 + Propeller stirrer |
| Detection range | | 1) Potentiometric | | : | | -2000mV to +2000mV | | 2) pH | | : | | -20.000 to 20.000pH | | 3) Temp. | | : | | 0 to 100°C | |
| Titration mode | Auto Titration , Auto Intermit , Intermit , Stat , Petroleum Titration , COD |
| Method | Standard method 120 , Combined method 10 (Max 5 methods can be linked) |
| Kinds of titration | Potentiometric (acid/base , redox , precipitation) , Photometric , Polarization , Conductivity |
| Titration form | Full titration (Auto EP detection) , EP Stop , Level Stop , Intersect , EP Stop/Level Stop |
| Special application | Measurement of electrode potential (pH , potential) , Acid dissociation constant (pKa) |
| Simultaneous recording of 2-way input potential (e.g. Titer vs. pH+%T , Titer vs. pH+μS) , Learn |
| Key operation | Touch panel |
| Displays | 1) 8.4-inch color LCD 800 x 600 dots |
| 2) English / Japanese / Mandarin Chinese / Korean / Russian / Spanish / German / French |
| 3) Simultaneous 4-channel display (Can also display Karl Fischer Moisture titrator simultaneously) | 3) 1-channel display |
| Calculation | Concentration of content , statistics data processing (mean , SD and RSD) and automatic averaging of blank and factor value |
| Data storage | 500 samples |
| GLP conformance | Registration of operator / User group administration | | Titrant | | : | | Reminder of date of factor measurement / Alarm to indicate remaining reagent / Reminder of piston replacement date / Reminder of reagent replacement date / History of factor measurement | | Check performance | | : | | Reminder of scheduled checkdate / Record of check results | | Management of electrode | | : | | Reminder of calibration date / Record of calibration history / Electrode check / History of electrode check | | Verification of burette capacity | | : | | Verification / Record of verification results | | Management of conduction time | | : | | Display of operating time | |
| Burette size | 20mL glass burette with brown cover (Standard) Optional burette units : 10mL , 5mL , or 1mL |
| Burette accuracy | 50mL burette(Auto dispenser) +/- 0.5mL 20mL burette +/- 0.02mL ; reproducibility +/- 0.01mL 10mL burette +/- 0.015mL ; reproducibility +/- 0.005mL 5mL burette +/- 0.01mL ; reproducibility +/- 0.003mL 1mL burette +/- 0.005mL ; reproducibility +/- 0.001mL |
| Preamplifier | | 1) STD | | : | | pH (mV) and mV , 2 inputs (Standard) | | 2) PTA | | : | | pH (mV) , mV and photometric , 3 inputs | | 3) POT | | : | | pH (mV) , mV and polar , 3 inputs | | 4) CMT | | : | | pH (mV) , mV and conductivity ,3 inputs (factory setting required) | | 5) TET | | : | | pH (mV) 2 ways and mV , inputs (factory setting required) | |
| External I/O | | RS-232C port x3 | | : | | for Dot matrix printer/Electronic balance/Data Capture Software (SOFT-CAP) | | SS-BUS x1 | | : | | for Multiple sample changer , APB | | ELE. x1 | | : | | for Smart electrode | | USB x1 | | : | | for USB flash drive , Thermal printer , A4 printer , Keyboard , Barcode reader , Foot switch , USB HUB | | TEMP.COMP. x1 | | : | | Input terminal for temperature sensor to correct reagent volume , sensor Pt100 , temperature reading accuracy : +/- 0.5°C (burette 1 only) | |
| LAN ×1 | | : | | for Personal computer (PC) | | - |
| Extensibility | Measuring instrument : Automatic Potentiometric Titrator (AT-710) , Karl Fischer Moisture Titrator (MKV-710/MKC-710); Three of these instruments can be added. | - |
| Automatic piston burette : Can control max 10 burette drives (Including two built-in burette drives) |
| Multiple sample changer : CHA-600 , CHA-700 |
| Ambient condition | | 1) Temperature | | : | | 5 to 35°C | | 2) Humidity | | : | | 85%RH or below (no condensation) | |
| Power source | AC100 to 240V +/-10% 50/60 Hz |
| Power consumption | | Main unit | | : | | Approx. 30W | | Printer | | : | | Approx. 7W | |
| Dimensions | | Touch panel controller | | : | | 225(W) x 190(D) x 42(H)mm | | Titration unit | | : | | 141(W) x 296(D) x 367(H)mm (not incl. tubing) | | Printer | | : | | 106(W) x 180(D) x 88(H) mm | |
| Weight | | Touch panel controller | | : | | Approx. 1.5kg | | Titration unit | | : | | Approx. 4.0kg | | Printer | | : | | Approx. 0.4kg | |
| Conformity standard | CE marking EMC : EN61326-1 LVD : EN61010-1 RE Directive Burette unit EBU FCC Part15 SubpartC FCC ID : 2ABSVEBU01 |
Clients Reviews